Abstract:In closed intranets or when external standard time sources are inaccessible, medical devices are prone to accumulated time deviations and reduced synchronization stability. This Research uses time information from local servers, databases, and terminal Real-Time Clock (RTC), and establishes a unified time reference through time source screening, weighted fusion, moving average filtering, and offset compensation. Device time synchronization is then performed through an intranet Network Time Protocol (NTP) service. Experimental results show that the multi-source fusion and compensation mechanism reduces the average synchronization error from 1.8 ms to 0.4 ms, decreases the standard deviation from 1.2 ms to 0.2 ms, and shortens the convergence time from 120 s to 60 s, while maintaining the time deviation within 0.3 ms to 0.5 ms during the stable stage. The method improves the time consistency and synchronization stability of medical devices under non-standard time source conditions.